Binder offers battery test safety packages for test chambers for aging, performance, and stress tests.
The packages can be adapted to a variety of Binder models in order to meet test requirements relating to temperature range, climate range, and test area size.
Package A for aging tests, EUCAR hazard level 2
- Class 2 independent temperature safety device when temperature is set to +120°C
- Temperature range limited to +110°C on the controller
Package P for performance tests, EUCAR hazard level 4
- Class 2 independent temperature safety device when temperature is set to +120°C
- Temperature range limited to +110°C on the controller
- Stainless-steel pressure relief flap installed in the middle on top of the unit
- Enhanced door-locking mechanism with reinforced brackets
- Inert gas connections
Package P Plus for extended performance tests, EUCAR hazard level 5
- Class 2 independent temperature safety device when temperature is set to +120°C
- Temperature range limited to +110°C on the controller
- Reversible pressure relief flap made of stainless steel installed in the middle on top of the unit
- Enhanced door-locking mechanism with reinforced brackets
- Inert gas connections
- Temperature-monitored testing area
- CO₂ fire suppression equipment
Package S for stress tests, EUCAR hazard level 6
- Class 2 independent temperature safety device when temperature is set to 120°C
- Temperature range limited to 110°C on the controller
- Reversible pressure relief flap made of stainless steel installed in the middle on top of the unit
- Enhanced door-locking mechanism with reinforced brackets
- N₂ permanent inertization
- Temperature-monitored testing area
- CO2 flushing device
- Monitoring of CO, H₂, and O₂ in testing area atmosphere
More information